Spectral Characteristics

This page contains 11 requirements.

Page summary

Metric

Value

Total requirements

11

Status: Compliant

10

Status: N/A

1

Verification: Test

8

Verification: Design

3

Result: PASS

10

Result: Very nearly hit on Nice to Have

1

Requirements

NGPS-SCI-REQ-0010 — The observable wavelengths of NGPS shall span 360 nm (330 nm goal) to 1000 nm (…

Field

Value

Identifier

NGPS-SCI-REQ-0010

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Feature

Yes

Status

Compliant

Verification method

Test

Result

PASS

Requirement

The observable wavelengths of NGPS shall span 360 nm (330 nm goal) to 1000 nm (1100 nm goal).

Notes / verification evidence

310-1050 achieved. arc data + domeflats

NGPS-SCI-REQ-0020 — There shall be at least one mode of operation where the entire wavelength range…

Field

Value

Identifier

NGPS-SCI-REQ-0020

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Design

Result

PASS

Requirement

There shall be at least one mode of operation where the entire wavelength range of 360 nm (330 nm goal) to 1000 nm (1100 nm goal) is observable simultaneously.

Notes / verification evidence

All 4 channels can operate concurrently

NGPS-SCI-REQ-0030 — The total system throughput (from sky to detected photons) at the wavelength of…

Field

Value

Identifier

NGPS-SCI-REQ-0030

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Test

Result

PASS

Requirement

The total system throughput (from sky to detected photons) at the wavelength of peak efficiency shall exceed 35%.

Notes / verification evidence

Obs. of std star 10” slit used. 45.5% at 743 nm (r-channel)

NGPS-SCI-REQ-0031 — The average system throughput in the range 370-500 nm shall exceed 27%.

Field

Value

Identifier

NGPS-SCI-REQ-0031

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Test

Result

PASS

Requirement

The average system throughput in the range 370-500 nm shall exceed 27%.

Notes / verification evidence

NGPS-SCI-REQ-0032 — The average system throughput in the range 500-750 nm shall exceed 31%.

Field

Value

Identifier

NGPS-SCI-REQ-0032

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Test

Result

PASS

Requirement

The average system throughput in the range 500-750 nm shall exceed 31%.

Notes / verification evidence

NGPS-SCI-REQ-0033 — The system throughput in the range 750-1000 nm shall exceed 30%.

Field

Value

Identifier

NGPS-SCI-REQ-0033

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Test

Result

PASS

Requirement

The system throughput in the range 750-1000 nm shall exceed 30%.

Notes / verification evidence

NGPS-SCI-REQ-0040 — The spectrograph shall deliver a resolving power with a 1.5 arcsec slit exceedi…

Field

Value

Identifier

NGPS-SCI-REQ-0040

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Test

Result

PASS

Requirement

The spectrograph shall deliver a resolving power with a 1.5 arcsec slit exceeding R = 1300 over the full observable range specified in NGPS-SCI-REQ-010.

Notes / verification evidence

CF: achieved, arcs+sky obs. measure R>1500

NGPS-SCI-REQ-0050 — Each channel of the spectrograph should have a mode capable of R > 4500

Field

Value

Identifier

NGPS-SCI-REQ-0050

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Nice to Have

Status

Compliant

Verification method

Test

Result

PASS

Requirement

Each channel of the spectrograph should have a mode capable of R > 4500

Notes / verification evidence

PASS using narrowest slice. Center slice is median R~4000 at 0.4” slit

NGPS-SCI-REQ-0060 — As a goal (not a requirement), the spectral range of the R > 4500 mode should c…

Field

Value

Identifier

NGPS-SCI-REQ-0060

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Nice to Have

Status

N/A

Verification method

Test

Result

Very nearly hit on Nice to Have

Requirement

As a goal (not a requirement), the spectral range of the R > 4500 mode should cover at least 1/3 of the channel.

Notes / verification evidence

I channel marginal at 0.4” slit

Actions

Would be good to take 0.37” arcs 1x1 binned at best focus for final confirmation

NGPS-SCI-REQ-0070 — As a goal (not a requirement), any wavelength in the channel should be availabl…

Field

Value

Identifier

NGPS-SCI-REQ-0070

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Nice to Have

Status

Compliant

Verification method

Design

Result

PASS

Requirement

As a goal (not a requirement), any wavelength in the channel should be available (e.g., through changing the angle of the grating) to the R > 4500 mode, though not necessarily in one exposure.

Notes / verification evidence

Gratings fixed in final design.

NGPS-SCI-REQ-0080 — The spectrograph shall offer a variety of slit widths ranging from 1 arcsec (0….

Field

Value

Identifier

NGPS-SCI-REQ-0080

Source

NGPS.SCI.DRD.001.docx

Source section

3.1

Category / request

Spectral Characteristics

Importance

Must Have

Status

Compliant

Verification method

Design

Result

PASS

Requirement

The spectrograph shall offer a variety of slit widths ranging from 1 arcsec (0.5 arcsec goal) to 5 arcsec (8 arcsec goal).

Notes / verification evidence

0.35 to 10” achieved